Terminated

Vacuum Regulator, Push-T-Set Intermittent Suction Unit (PTS-ISU) Product Usage: The vacuum regulator is intended to be used in a medical facility as a means to evacuate media (i.e. fluids) from the body.

FDA Recall Z-2410-2018 · Posted None

Recall StatusTerminated
Product CodeKDP
Date Initiated2018-03-12
Date Terminated2020-07-16
Product Quantity11 units

Reason for Recall

Final quality control testing was not completed before the devices were distributed..

Firm Action

Ohio Medical notified customers via phone call on about 03/12/2018. The notification included identification of the devices distributed to that customer, explanation that due to the need to the missed additional testing they were removing the devices, and providing a replacement until the testing can be completed.

Root Cause

Employee error

Distribution Pattern

US Nationwide Distribution in the states of NC, IN, TN, LA,, FL MI, IL.