Terminated

Octopus 101 Perimeter; PeriTrend Software Analysis A perimeter is a device intended to determine the extent of the peripheral visual field of a patient. The device projects light on various points of a curved surface, and the patient indicates whether he or she sees the light.

FDA Recall Z-2411-2012 · Posted 2012-09-20

Recalling FirmHaag-Streit USA Inc
Recall StatusTerminated
Product CodeHPT
Date Initiated2012-08-17
Date Terminated2015-05-29
Product Quantity64 units

Reason for Recall

The firm recalled the device after learning of a possible deviation in the stimulus luminance of the Perimeter OCTOPUS 101.

Firm Action

The firm, HAAG-STREIT USA, Inc., sent a "Field Safety Notice 2012-01/01" dated August 10, 2012 via certified mail or onsite visit to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to make all users of the product and relevant personnel aware of this Field Safety Notice; forward a copy to third parties, if affected product has been passed on; keep this notice at least until the completion of the corrective measure, and complete and return the enclosed Confirmation of Receipt Form via fax to: 513-229-3867 or scanned and emailed to : [email protected]. If you have any questions, call 513-336-7255.

Root Cause

Equipment maintenance

Distribution Pattern

Worldwide distribution: USA (nationwide) including states of: AR, CA, CT, FL, GA, IA, IL, LA, MA, MI, MN, MS, NJ, NV, NY, OH, OR, PA, SC, SD, WA, WI, and WV; and to the countries of Jamaica, Puerto Rico, and Mexico.