8 G x 10 CM Jamshidi Crown Bone Marrow biopsy/Aspiration Needle with MAC Catalog Number: TMJ4008
FDA Recall Z-2411-2019 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeKNW
Date Initiated2019-08-01
Date Terminated2020-05-06
Product Quantity700 units
Reason for Recall
Product Pouches may not be sealed properly and compromise the sterility of the device
Firm Action
BD notified US consignees via FedEx August 1, 2019,and contacted via email. Sub recall requested and BD will notify customers. BD is requesting that customers destroy product at their location. The kit packer asked to return product. BD will mail additional communications to the kit customers.
A business response form is included with the letter to assist direct consignees in responding to the field action.
Questions contact BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States.
Root Cause
Packaging process control
Distribution Pattern
CA, IL Foreign: Taiwan
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.