Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

FDA Recall Z-2413-2024 · Posted 2024-08-12

Recalling FirmPalodex Group Oy
Recall StatusOpen, Classified
Product CodeOAS
Date Initiated2021-10-28
Product Quantity103

Reason for Recall

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

Firm Action

Palodex Group Oy is preparing a Class IlIl field correction to update the firmware of ORTHOPANTOMOGRAPH OP 3D Pro of the 27 defective units and these devices will be retested for Pan ADC reproducibility. All affected persons will be notified of this defect in accordance with 21 CFR 806.10.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide-Worldwide