S-Rom Total Hip System Poly Liner Packaging: The insert is placed into an inner foil pouch and vacuum sealed; a ring is placed into a Tyvek/PET pouch and sealed; the foil pouch and Tyvek/PET pouch are placed into a nylon outer pouch and sealed. This is then placed into a folding carton, labeled, and shrink-wrapped. Description: Total Hip System Poly Dial Liner is part of the S-ROM Total Hip System. The Liner is fitted inside the acetabular shell; and is the bearing surface for the Head on the S-ROM stem. The S-ROM liner is used for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Intended Use: Total Hip System Poly Dial Liner is part of the S-Rom Total Hip System. The Liner is fitted inside the acetabular shell; and is the bearing surface for the Head on the S-Rom stem. The S Rom liner is used for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

FDA Recall Z-2414-2012 · Posted 2012-09-20

Recall StatusTerminated
Product CodeKWZ
Date Initiated2012-08-22
Date Terminated2013-10-09
Product Quantity12 liners

Reason for Recall

Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of four lots of the S-ROM" liner because the Carton label has an Incorrect Color Code Series Designation. We have received one complaint regarding this issue.

Firm Action

DePuy sent an "URGENT INFORMATION-MEDICAL DEVICE RECALL NOTICE" dated August 22, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Reconciliation Form was included in the letter for customers to complete and return via fax to 574-371-4939. Contact your local Depuy Orthopaedics sales representative for questions related to this recall.

Root Cause

Labeling Change Control

Distribution Pattern

Nationwide Distribution including the states of MA, TX, OH, CA, and NC.