SlideView VS200 slide viewer,

FDA Recall Z-2416-2024 · Posted 2024-07-26

Recall StatusOpen, Classified
Product CodeRIB
Date Initiated2024-07-22

Reason for Recall

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Root Cause

Radiation Control for Health and Safety Act