Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call station.

FDA Recall Z-2418-2024 · Posted 2024-07-23

Recall StatusOpen, Classified
Product CodeIQA
Date Initiated2024-06-20
Product Quantity77 units

Reason for Recall

This is a risk that alarms will not be received by the Arial Wireless Nurse Call System and premature depletion of the battery.

Firm Action

Securitas Healthcare issued a recall notice on 06/19/2024 to its consignees via USPS certified mail. The notice explained the problem, potential hazard, and requested the following: "Customer Required Actions: 1. Closely Monitor Missing Device Alerts within the Arial Application and Immediately Verify Status of Missing Devices and Cease Using any Call Stations That Do Not Reconnect. Clients should closely monitor the "Missing Device Alerts within the Arial Application and immediately check the status of Call Stations that trigger a Missing Device Alert and replace the Call Station battery if it is low or fully depleted. Instructions for replacing the Call Station battery are attached hereto as Appendix 1. Instructions for reconnecting missing devices and confirming that the Call Stations are properly transmitting information to the Arial Emergency and Nurse Call System are attached hereto as Appendix 2. Clients should immediately cease using any Call Station that will not reconnect to the Arial Emergency and Nurse Call System. 2. Closely Monitor the Battery Status of all Call Stations in the Arial Web Client and Replace Battery if it is Low . Clients should closely monitor the Call Station Battery Status in the Arial Web Client and immediately replace the battery if it is low . 3. Consider Use of Supplementary Communication Devices or Procedures for High Risk Residents. Clients should consider enacting backup procedures, including increased room checks, or providing alternate emergency and nurse call communication means, such as a wearable pendant, for high-risk residents." "If you have distributed the product further, immediately notify your accounts that received the product identified above of this notification and ask them to contact Securitas Technical Support at 1-800-824-2996 (8:00 AM 5:00 PM CT, M-F) to obtain a response form and coordinate correction of affected devices." Securitas Healthcare is developing a firmware patch to remediate t

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

US: FL, CA, AL, ND, MS, WA, MN, TX, OR, WI, MD, NY, NE

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑