Terminated

IBC Suction Wand Catalog Number 1990S Product Usage: The IBC Suction Wand is a component of Blood Recovery System for use in open heart procedures, autologous blood recovery and blood suction. It is a disposable, single use component.

FDA Recall Z-2422-2015 · Posted 2015-08-19

Recall StatusTerminated
Product CodeDTS
Date Initiated2015-08-07
Date Terminated2016-03-01
Product Quantity100 units

Reason for Recall

It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.

Firm Action

The recalling firm began notifying their affacted customers on 7/2/2015

Root Cause

Packaging process control

Distribution Pattern

Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.