Terminated

NeoMarkers Rabbit Monoclonal anti-Human Cyclin D1 Antibody (Clone SP4), immunochemistry Reagent and Kit, For In Vitro Diagnostic use, manufactured by Lab Vision Products, Fremont, CA, a subsidiary of ThermoFisher Scientific

FDA Recall Z-2424-2010 · Posted 2010-09-15

Recall StatusTerminated
Product CodeNJT
Date Initiated2010-03-05
Date Terminated2011-03-21
Product Quantity103 units

Reason for Recall

Two lots may be contaminated with Ki-67 antibody, which could cause abnormal staining patterns, and produce false positive results.

Firm Action

Thermo Fisher Scientific issued an "Advisory Notice" letter dated March 5, 2010 to all domestic customers and international distributors. The letter described the product, problem and action to be taken. Customers were instructed to: 1) discard the affected lots, and 2) retest specimens using a replacement lot supplied by the firm, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 International distributors were instructed to: 1) place the affected lots and replace with a replacement lot supplied by the firm, and 2) provide customers with copies of the customer letter, and 3) complete and return the Confirmation of Receipt Form via fax to 1-510-979-5489 For questions, contact Thermo Fisher Scientific at 1-800-828-1628 or 1-510-771-1595.

Root Cause

Material/Component Contamination

Distribution Pattern

Worldwide Distribution: USA, including the states of CA, SC, IL NY, WI, PA, MN, AL, WA, MA, and KS, and the coutnries of Austria, Australia, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Malaysia, the Netherlands, Norway, Pakistan, Poland, Portugal, Russia, the UK, and Turkey.