IBC Vascular Loop Catalog Number CH111, CH112, CH113, CH114, CH115, CH116, CH118 Product Usage: The IBC Vascular Loop is a single-use, disposable loop used in occluding, retracting, and identifying arteries, veins, tendons, and nerves in surgical procedures.

FDA Recall Z-2424-2015 · Posted 2015-08-19

Recall StatusTerminated
Product CodeDWS
Date Initiated2015-08-07
Date Terminated2016-03-01
Product Quantity771 units

Reason for Recall

It was discovered that the heat seal placed onto sterile pouches may not meet strength acceptance criteria.

Firm Action

The recalling firm began notifying their affacted customers on 7/2/2015

Root Cause

Packaging process control

Distribution Pattern

Worldwide - United States Nationwide and the countries of: Germany, Norway, Turkey, Chile, Pakistan, and Republic of South Africa