Coronary Sinus (CS) Uni-Directional Diagnostic Electrophysiology (EP) Catheter, catalog numbers D135303 and D135304. Product Usage: The Reprocessed CS Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures; i.e., stimulation and recording only. The catheter is designed for use in the coronary sinus.

FDA Recall Z-2426-2018 · Posted None

Recall StatusTerminated
Product CodeNLH
Date Initiated2018-03-29
Date Terminated2020-07-28
Product Quantity2,722 units

Reason for Recall

Thrombogenicity test results demonstrated that both the test devices (SSS) and control devices (OM) did not meet the internal specification of non-thrombogenic .

Firm Action

The recalling firm issued a notification letter to their sales representatives via email on 3/29/2018 with a response form. For customers who did not return the response form, their will be a subsequent email or fax to their sales rep. The company will continue to follow up via phone and email as needed. The letter identified the affected product, problem and actions to be taken.

Root Cause

Use error

Distribution Pattern

Worldwide Distribution - US Nationwide and Canada

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.