Kallestad (TM) HEp-2 Cell Line Substrate. Product labeled "BIO-RAD Kallestad (TM) HEp-2 Cell Line Substrate... UNITED STATES, Bio-Rad Laboratories, Redmond, WA 98052... FRANCE, Bio-Rad Laboratories, 92430 Marnes La Coquette". Catalog #: 30471 - 12 well kit, 60 tests; 30472 - 12 well kit, 240 tests; and 32583 - 6 well kit, 48 tests.

FDA Recall Z-2428-2008 · Posted 2008-09-19

Recalling FirmBio-Rad Laboratories
Recall StatusTerminated
Product CodeDBM
Date Initiated2008-07-29
Date Terminated2011-03-31
Product Quantity4,400 total kits

Reason for Recall

The Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128 reported on the QC Report Label included with each test kit.

Firm Action

On 7/25/08, the firm began to notify their customers about the recall via an Urgent Product Correction sent by DHL or FEDEX. Included with the customer letter was a revised QC Card. International subsidiaries were notified by e-mail with the customer letter and revised QC Card attached. The letter informed customers of the recall and directing them to the revised QC card which was attached to the letter. The instructions are to, "Please use this to replace the card in any kits remaining in your facility." If customers have any questions or require additional QC cards, they are to contact Technical Support.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution --- including USA and countries of AUSTRALIA, BRAZIL, CANADA, CHINA , FRANCE, GERMANY, HONG KONG, INDIA, ITALY, JAPAN, MEXICO, NEW ZEALAND, POLAND, SINGAPORE, SOUTH KOREA , TAIWAN, UNITED KINGDOM, and URUGUAY.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →