Terminated

Recon, Radial Implant, Standard Stem, Size 1, Sterile, Do Not Reuse, Rx Only. Intended for replacement of the proximal end of the radius.

FDA Recall Z-2429-2015 · Posted 2015-08-20

Recall StatusTerminated
Product CodeJDB
Date Initiated2015-06-24
Date Terminated2016-04-29
Product Quantity16992 total

Reason for Recall

Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.

Firm Action

Branches/Agencies were notified by Stryker of this action by e-mails on June 24, 2015 and an Urgent Medical Device Removal Letter/Acknowledgement Response Form dated June 25, 2015 was sent to the attention of the Risk Manager.

Root Cause

Packaging change control

Distribution Pattern

Worldwide Distribution.