Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F

FDA Recall Z-2429-2026 · Posted 2026-06-12

Recall StatusOpen, Classified
Product CodeMLS
Date Initiated2026-04-27
Product Quantity49,654 kits total

Reason for Recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Firm Action

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution. OUS distribution pending.