Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481, 300100483, 300100613, 300101013, 300101113, 300101114, 300101313, 300101379, 300102213, 300102713, 300102913, 300103413, 300103414, 300104213, 300105413, 300105713, 300106313, 300106513, 300106581, 300107813, 300107814, 300108513, 300108813, 300108814, 300109913; infusion pump
FDA Recall Z-2430-2023 · Posted 2023-08-18
Advertisement
Recalling FirmICU Medical Inc
Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2023-07-13
Product Quantity93135 devices
Reason for Recall
Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.
Firm Action
ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 07/13/2023 via traceable mailing. The notice explained the issue, potential risk, and requested the following:
Actions for Clinical Users:
If your Plum 360 pump does not power on or indicates an Audio Alarm failure, please remove the pump from use and send the pump to Biomedical Engineering.
Actions for Biomedical Engineering:
If you have Plum 360 pumps with Audio Alarm failures, please contact ICU Medical.
1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and required actions.
2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification.
3. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected].
A similar notice was distributed specifically to consignees of instruments that were serviced which lists the specific serial numbers. The issue, potential risk, and actions are the same as the general notice.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.