Medline convenience kit: SMSC-PELVISCOPY PACK-LF DYNJ0161619M
FDA Recall Z-2430-2026 · Posted 2026-06-12
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeMLT
Date Initiated2026-04-27
Product Quantity49,654 kits total
Reason for Recall
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Firm Action
Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution. OUS distribution pending.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement