Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity in human plasma. Catalog Number: 101201 For the determination of Plasminogen Activator Inhibitor Type 1 (PAI-1) activity in human plasma. The test is for in Citro diagnostic use.
FDA Recall Z-2431-2012 · Posted 2012-09-20
Advertisement
Recalling FirmSekisui Diagnostics Llc
Recall StatusTerminated
Product CodeGGP
Date Initiated2012-08-28
Date Terminated2016-07-25
Product Quantity699 kits
Reason for Recall
SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make more frequent sample dilutions.
Firm Action
The firm, Sekisui Diagnostics, LLC, sent an "URGENT PRODUCTS RECALL NOTICE" dated August 28, 2012 to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to immediately stop using the product; to discard any remaining inventory and complete and return the Customer Acknowledgement Form by September 4, 2012 documenting the # of kits destroyed via FAX to Quality Assurance at 203-602-5553. The customers were also instructed to return the form even if they do not have the affected product currently in inventory. Sekisui will replace kits.
If you have any questions contact, Customer Service Department at 800-637-3375, ext 10, 12, or 31, or email to recall coordinator at [email protected].
Root Cause
Process control
Distribution Pattern
Nationwide distribution: USA including states of: FL, IN, IL, and WA.