PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and health physics.
FDA Recall Z-2439-2015 · Posted 2015-08-21
Advertisement
Recalling FirmPTW-Freiburg
Recall StatusTerminated
Product CodeIYE
Date Initiated2015-05-27
Date Terminated2017-03-31
Product Quantity242 firmwares
Reason for Recall
In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.
Firm Action
PTW issues Urgent Safety Notice due to an error with firmware version 2.05 of the UNIDOS webline in some cases, especially with the high dose-per-pulse-rates typical for particle therapy, measuring errors can result.
Root Cause
Software design
Distribution Pattern
Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.