PTW UNIDOS webline: 1) REF T10023, 2) REF T10022 and 3) REF T10023 Firmware version 2.05. Dosimeter for radiation therapy, diagnostic radiology and health physics.

FDA Recall Z-2439-2015 · Posted 2015-08-21

Recalling FirmPTW-Freiburg
Recall StatusTerminated
Product CodeIYE
Date Initiated2015-05-27
Date Terminated2017-03-31
Product Quantity242 firmwares

Reason for Recall

In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low dose being displayed on the monitor. This could result in a radiation over-exposure.

Firm Action

PTW issues Urgent Safety Notice due to an error with firmware version 2.05 of the UNIDOS webline in some cases, especially with the high dose-per-pulse-rates typical for particle therapy, measuring errors can result.

Root Cause

Software design

Distribution Pattern

Distributed in CA, MI, MA, NC, NJ, NY, PA, TN, TX, PA & WI.