BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213
FDA Recall Z-2439-2019 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Co.
Recall StatusTerminated
Product CodeJTC
Date Initiated2019-07-26
Date Terminated2019-12-06
Product Quantity9 units
Reason for Recall
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
Firm Action
BD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation.
Contact: BD Customer/Technical Support 800-638-8663
Monday Friday between 7:00am and 7:00pm (EST) in the United States.
For customers outside the US, contact your local BD representative or distributor.
Root Cause
Software design
Distribution Pattern
Illinois, Iowa, Missouri, New York, Ohio, Tennessee World Wide (5): Singapore, Taiwan, France, Netherlands
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.