Solero Generator PG, Item No. H78712740000, Catalog No. 12740000

FDA Recall Z-2439-2021 · Posted None

Recalling FirmAngiodynamics, Inc.
Recall StatusTerminated
Product CodeNEY
Date Initiated2021-07-20
Date Terminated2023-03-15
Product Quantity65 OUS

Reason for Recall

Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up.

Firm Action

On July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees. Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices. In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere to the following Warning in the current product labeling (User Manual): WARNING: Do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "Ready".

Root Cause

Software design

Distribution Pattern

Domestic distribution Nationwide. Foreign distribution worldwide.