Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
FDA Recall Z-2441-2018 · Posted None
Advertisement
Recalling FirmZeiss, Carl Inc
Recall StatusTerminated
Product CodeRGW
Date Initiated2018-05-29
Date Terminated2024-07-26
Product Quantity8
Reason for Recall
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Firm Action
The Company will immediately notify by direct contact (when available) and/or express courier each first line purchaser and/or known user site with informing them of the risk, how to prevent it and the notice of action.
The Company will dispatch factory trained service representatives to perform following actions:
o Installation of Hot Fix # 14 (or newer) for ZEN 2.3
o Installation of Firmware 1 .209 (or newer) on nodes x l 5 and x l 9 (Main Controller 2)
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.