SERVO-i Ventilator System Maquet Critical Care, Solna, Sweden. The product is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency.

FDA Recall Z-2442-2008 · Posted 2008-09-23

Recalling FirmMAQUET Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2008-07-08
Date Terminated2008-12-23
Product Quantity122 units (SERVO-i and SERVO-s) units in United States

Reason for Recall

Maquet became aware of potentially defective crimpings at the point connector attached to the cable in SERVO-i and SERVO-s ventilators.

Firm Action

Urgent - Device Correction notices were sent out to all customers on July 7, 2008. The letter informed the customers of the affected product, and associated hazards. The letter instructed customers to take an inventory of the product and it provided actions to be taken until affected products are replaced. Contact Maquet at 1-888-627-8383#1, for assistance.

Root Cause

Employee error

Distribution Pattern

Worldwide.