Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

LAPAROSCOPIA COLORECTAL PACK CONTENTS: (1) SOLUTION SURGICAL DURAPREP 26ml (4) DRAPE UTILITY WITH TAPE (1) TOWELS ABSORBENT 15" X 20" (2) LEGGINS W/7" CUFF 30" X 42" (1) BLADE SURGICAL #15 STAINLESS STEEL (1) BLADE SURGICAL #11 STAINLESS STEEL (1) GOWN IMPERV. EXTRA REINFORCED XLG (1) GOWN LGE STANDARD SMS VELCRO NECK (1) IRRIGATION SUCTION TROMPET DUAL SPIKE (10) GAUZE SPONGE 4" X 4" 16PLY XRD (2) UTILITY BOWL 16oz (2) STERI STRIP CLOSURE W' X 4" (1) ACTIFOG W/FOAM PAD SOLUTION (1) TABLE COVER REINFORCED 50" X 90" (1) MAYO STAND COVER REINFORCED (1) NEEDLE ULTRA VERES 120MM (1) STAPLE ARTICULATING ECHEL 60MM (1) RELOAD WHITE 60MM ECHELON (1) RELOAD BLUE 60MM ECHELON (2) CANNULA 5MM XCEL SLEEVES (1) TROCAR 5MM BLADELE WITH STAB SLEEVE (1) TROCAR 12MM BLADELE WITH STAB SLEEVE (1) DRAPE LAP. ABDO W/POUCH 102" X 122" X 78" STD SMS Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

FDA Recall Z-2444-2014 · Posted 2014-08-29

Recalling FirmCustomed, Inc
Recall StatusTerminated
Product CodeFDE
Date Initiated2014-05-20
Date Terminated2017-08-14
Product Quantity17 lots/626 units,(multiple units per lot)

Reason for Recall

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Firm Action

Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.