Open, Classified

Medline convenience kits: BIB AV FISTULA PACK DYNJ58304B CARDIAC MINOR SURGICOUNT PACK DYNJ56573B ENDOVASCULAR PACK DYNJ29135N EP DEVICE DYNJ65916B FULL LINE ANEURYSM CDS CDS983672P GEN FEM POP #14-RF DYNJ21877W IMPLANT PACK DYNJ39316D OPEN HEART MRNGSD DYNJ58354F PACEMAKER PACK-LF DYNJ0618174N PACEMAKER SETUP PACK DYNJ35761D PACER / CATH LAB PACK-LF DYNJ0972604A PACER CDS CDS983686I PACER PACK DYNJ52059D PK, EP LAB PACEMAKER DYNJ43663A VENOUS ACCESS PACK DYNJ45640B

FDA Recall Z-2444-2026 · Posted 2026-06-12

Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2026-04-27
Product Quantity49,654 kits total

Reason for Recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Firm Action

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution. OUS distribution pending.

Official Source

View FDA Recall →