Terminated

Universal Optical Tracker Fixation, not sterile, Manufactured by Orthosoft, Montreal, Canada, REF 111.031. Surgical instrument.

FDA Recall Z-2445-2008 · Posted 2008-09-20

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2007-08-06
Date Terminated2009-03-11
Product Quantity39

Reason for Recall

Inadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection.

Firm Action

Firm's sales representatives were notified via e-mail on 8/6/07 to review their inventory and that of their accounts and to notify the firm of their location so that they could be replaced. Contact the firm by telephone at 574-267-6131 for additional information concerning this recall.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution including states of California, Florida and Washington.