Sterile Disposable Connectors ASY 1/4 X 3/8 REDUCER W/LL 24. REF/ GTIN for Insert Label:050434000/ 00803622123207

FDA Recall Z-2445-2020 · Posted None

Recalling FirmLivaNova USA Inc.
Recall StatusTerminated
Product CodeDTL
Date Initiated2020-04-24
Date Terminated2022-02-17
Product Quantity24 units

Reason for Recall

The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.

Firm Action

The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.

Root Cause

Packaging process control

Distribution Pattern

US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)