Terminated

Dimension Vista B2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

FDA Recall Z-2450-2015 · Posted 2015-08-22

Recall StatusTerminated
Product CodeJZG
Date Initiated2015-07-23
Date Terminated2017-03-24
Product QuantityTotal 1655 units

Reason for Recall

customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.

Firm Action

Siemens sent an Urgent Medical Device Correction letter dated July 2015 to all affected customers. The letter identified the affected product, problem, risk to health, and actions to be taken by customers. For questions contact your Customer Care Center - Technical Solutions at 1-800-441-9250 or your local Siemens technical support representative.

Root Cause

Environmental control

Distribution Pattern

US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.