TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

FDA Recall Z-2456-2026 · Posted 2026-06-16

Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeHSB
Date Initiated2026-03-16
Product Quantity5

Reason for Recall

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

Firm Action

On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700

Root Cause

Process control

Distribution Pattern

International distribution to the countries of BE, DE, IT.