Varian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithim is specifically indicated for planning proton treatment of neoplasms of the eyes.
FDA Recall Z-2462-2012 · Posted 2012-09-26
Advertisement
Recalling FirmVarian Medical Systems, Inc. Oncology Systems
Recall StatusTerminated
Product CodeMUJ
Date Initiated2012-08-17
Date Terminated2014-08-06
Product Quantity6087
Reason for Recall
Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.
Firm Action
VARIAN Medical Systems sent an Urgent Medical Device Correction Urgent Field Safety Notice on August 17, 2012, to all affected customers with a description of the problem and user corrective action steps. Customers were instructed to ensure that the workstation clock be synchronized with the DatabaseSystem server clock. Customers were also instructed to contact Varian service for assistance with time synchronization if required. Customers were to complete the attached Proof of Notification Form once they have read the document and return it to Varian Medical Systems.
For questions customers should call the following numbers:
USA and Canada 1-888-827-4265.
Europe +41 41 749 8844.
For questions regarding this recall call 650-424-5731.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally. . .
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.