Terminated

Tiger Cannulated Tray Lid Part # 210-00-001-1 NON-STERILE; Combined Tiger/Headless Screw Part # 210-99-101 NON-STERILE; These are components of the Tiger Cannulated Screw System and the Tiger Headless Cannulated Screw System The Tiger Cannulated Screw Fixation System and the Tiger Headless Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.

FDA Recall Z-2464-2015 · Posted 2015-08-26

Recall StatusTerminated
Product CodeHWC
Date Initiated2015-07-20
Date Terminated2017-01-26
Product Quantity71 units

Reason for Recall

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Firm Action

Trilliant sent an Important Medical Device Adivsory Notice dated November 6, 2013, to all affected customers. The letter instructed customers that they have the option of exchanging affected instruments for trays that meet specification at no charge. To initiate the exchange, customers were instructed to contact their local Trilliant Surgical sales representative or the Customer Service Department directly at 1-800-495-2919, or via email at [email protected].

Root Cause

Component design/selection

Distribution Pattern

Nationwide Distribution