EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V
FDA Recall Z-2467-2023 · Posted 2023-08-28
Recalling FirmOlympus Corporation of the Americas
Recall StatusOpen, Classified
Product CodeFDT
Date Initiated2023-07-27
Product Quantity6,426 units
Reason for Recall
Recent reports of patient infection.
Firm Action
Consignees were sent an Urgent Medical Corrective Action notification dated 7/27/2023. This action is being initiated due to the firm becoming aware of and FDA inquiring about recent reports of infections and positive cultures. Olympus is reminding users to closely follow reprocessing instructions, including periodic inspections by Olympus Service, and informing users about an updated manual intended to improve comprehension of existing reprocessing steps. Additional enhancements to the manual include the use of reprocessing timelines and simplified text/visualization techniques. The manual is available on the customer web portal and paper copies, along with wall charts, are available to be mailed to your facility.
The notification also emphasises the required inspection of the forceps elevator mechanism, to be performed once a year or every 100 reprocessing cycles, and encourages consignees to consider incorporating voluntary routine microbiologic surveillance of endoscops as part of their quality improvement/assurance efforts.
Consignees are asked to complete and return the provided response form to Sedgwick via email at [email protected] or by fax to 888-208-4588. Questions can be directed by phone to 866-912-9552. Customers can also contact Ashley Mitch at Olympus by phone at 484-553-1029 or by email at [email protected] with any additional concerns.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution.