Dideco, D901 Lilliput 1, Close Phisio, Sterile EO, REF 05252, Manufacturer: Sorin Group Italia, Via Statale 12 Nord, 86, 41037 Mirandola (MO) (Italy). The product is intended for use in cardiopulmonary bypass circuits as a device to replace the function of the lungs in order to control the arterial/venous temperature and as a venous blood reservoir
FDA Recall Z-2469-2008 · Posted 2008-09-30
Advertisement
Recalling FirmSorin Group USA, Inc.
Recall StatusTerminated
Product CodeDTZ
Date Initiated2008-08-08
Date Terminated2008-09-30
Product Quantity35 units
Reason for Recall
Under certain conditions, venous reservoir on oxygenator may allow air into the system.
Firm Action
The firm sent out a notification letter, by certified letter, on Aug 8,2008. Instructions on how to prevent air from entering the system were provided. Sorin Group USA, Inc. can reached at 1-303-467-6306.
Root Cause
Labeling False and Misleading
Distribution Pattern
NC