Terminated

Synthes Application Instrument for Sternal ZIPFIX; indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

FDA Recall Z-2475-2015 · Posted 2015-08-27

Recall StatusTerminated
Product CodeJDQ
Date Initiated2015-07-22
Date Terminated2016-08-18
Product Quantity451

Reason for Recall

The end cap may loosen and detach making the instrument non-functional. No injuries reported.

Firm Action

Urgent recall notification letters, dated July 22, 2015 were sent to end users and consultants to inform them of the issue, the associated risks, and provided instructions to return affected devices along with the response form.

Root Cause

Device Design

Distribution Pattern

Nationwide and internationally to Canada.