LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,

FDA Recall Z-2475-2023 · Posted 2023-08-28

Recalling FirmAlphatec Spine, Inc.
Recall StatusOpen, Classified
Product CodeHWJ
Date Initiated2023-07-07
Product Quantity29

Reason for Recall

Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.

Firm Action

On 7/7/23, recall notices were emailed to consignees who were asked to do the following: 1) Abstain from sale and/or use and return to the recalling firm. 2) Complete and return the response form via email to [email protected] Questions/concerns can be directed to 1-800-922-1356 or [email protected]

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.