Maestro Orthodontic Bracket REF711-298Q QTY. 20 Rx Only (01)00190707017064 (10)889437 (422)484 Maestro Orthodontic Bracket REF711-299MCQ QTY. 20 Rx Only (01)00190707017071(10)889437 (422)484

FDA Recall Z-2478-2020 · Posted None

Recall StatusTerminated
Product CodeEJF
Date Initiated2019-04-02
Date Terminated2022-03-24
Product Quantity147 kits

Reason for Recall

Orthodontic Bracket is mislabeled

Firm Action

The firm notified their consignees of the recall by letter sent out 1st class mail on 4/2/2019. The recall was initiated due to a mislabeling error. The Upper Right bracket was erroneously color coded as an Upper Left brackets and then comingled within the same bag. The firm is asking recipients of affected product to please contact Customer Service at 1-800-547-2000 to receive an RMA number. The RMA will allow for a quick return and replacement or credit. The firm is requesting that distributors and physicians complete and E-mail the acknowledgment form back to the firm in order to confirm receipt of the field correction letter. Additionally they are asking authorized distributor,to identify those customers that may have been shipped the affected product and contact those customers to inform them of the issue within fortyeight (48) hours of receipt of this field correction notification.

Root Cause

Process control

Distribution Pattern

13 US Consignees in FL, GA, CA, MI, ID, TX 1 = Italy 1 = Japan 1 = Greece 1 = Saudi Arabia 1 = Russia 1 = Czech Republic