Recall StatusOpen, Classified
Date Initiated2023-07-27
Product Quantity92
Reason for Recall
Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
Firm Action
On 7/27/23, recall notices were sent to customers and distributors who were asked to do the following:
1) Stop using and work with your firm representative to return affected devices.
2) Share this notice with the appropriate clinical staff at your site.
3) Complete and return the customer acknowledgement form.
4) Distributors: Notify your customers by sending this customer notification to any of your customers who have purchased the impacted devices.
Root Cause
Process change control
Distribution Pattern
International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.