Terminated

COULTER ISOTON 4 Diluent, Catalog No. 8547148. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types, such as red blood cells and platelets.

FDA Recall Z-2481-2015 · Posted 2015-08-28

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeGIF
Date Initiated2015-07-22
Date Terminated2015-12-18
Product Quantity710 units

Reason for Recall

Beckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.

Firm Action

An Urgent Medical Device Recall letter dated 7/22/15 was sent to the affected customers to inform them of the issue and actions to take. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Beckman Coulter's Customer Support Center, http://www.beckmancoulter.com/customersupport/support, (800) 526-7694 in US and Canada. Customers outside of the US and Canada, are instructed to contact their local Beckman Coulter Representative.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distributed US (nationwide) including Puerto Rico and U.S. Virgin Islands and the countries of Bermuda, Canada, Guatemala, Hong Kong, Mexico, Panama, Taiwan, and Trinidad-Tobago.