Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01

FDA Recall Z-2481-2023 · Posted 2023-08-29

Recall StatusOpen, Classified
Product CodeLXH
Date Initiated2023-07-26
Product Quantity6 units

Reason for Recall

A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery

Firm Action

LinkBio Corp.(US Importer) issued Urgent: Medical Device Recall letter via email on July 27, 203 to US Distributors . Letter states reason for recall, health risk and action to take: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Root Cause

Device Design

Distribution Pattern

IN,GA, TX