ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.

FDA Recall Z-2482-2010 · Posted 2010-09-23

Recall StatusTerminated
Product CodeJJE
Date Initiated2008-10-21
Date Terminated2012-04-13
Product Quantity404 units distributed worldwide (including US)

Reason for Recall

During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover.

Firm Action

ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit .

Root Cause

Mixed-up of materials/components

Distribution Pattern

Worldwide distribution:USA including states of: CO, DC, FL, GA, IL, LA, MA, MN, NY, OH, OK, TX, UT, and WV and countries including: Australia, Brazil, Bogota, Canada, China WFOTC, Columbia, Curacao (NETH ANT), Germany, Hong Kong, Japan, Mexico, Singapore, and Taiwan.