Terminated

HS Hi-R NEO 900A Reference # 657 821 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.

FDA Recall Z-2483-2019 · Posted None

Recalling FirmHaag-Streit USA Inc
Recall StatusTerminated
Product CodeEPT
Date Initiated2019-07-31
Date Terminated2020-04-16
Product Quantity0

Reason for Recall

software error in the central control unit of the floor stand.

Firm Action

1. Do not perform any other functions at the same time when operation the XY control, such as changes in the light intensity. 2. Perform a software update of the floor stand. you will receive the software version 11.4 through your local representative which will fix the bug. 3. please send us the enclosed response form filled In as soon as possible. 4. If you have passed the products to third parties, please forward a copy of this information and inform the below mentioned contact person. The software update will be organized through Haag-Streit USA. For questions concerning this procedure, please contact: Brian Gallagher Phone: (513) 486-1522 Email: fieldsafetynotice@haag-streit-usa .com

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

US in the state of OH

Official Source

View FDA Recall →