Terminated

This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.

FDA Recall Z-2490-2019 · Posted None

Recall StatusTerminated
Product CodeJIX
Date Initiated2019-07-12
Date Terminated2020-06-17
Product Quantity307 kits

Reason for Recall

Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).

Firm Action

Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to [email protected] within five working days.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Foreign: Antigua, Aruba, Bahamas, Granada, Tobago,