Blood gases (PCO2, PO2) and blood pH test system; ABL800 FLEX; Model Number 393-801; UDI (01)05700693938011. Model 393-801 is available as ABL837. The FlexQ module is optional and can be added to all models and configurations.

FDA Recall Z-2494-2020 · Posted None

Recall StatusTerminated
Product CodeCHL
Date Initiated2020-05-26
Date Terminated2023-02-08
Product Quantity3,899 units

Reason for Recall

Potential risk of patient mix-up on analyzers due to software issues.

Firm Action

On 05/26/2020, the firm sent an Urgent: Medical Device Recall letter via email to all consignees; additionally the firm will send a hard copy letter FedEx 2nd day with tracking and proof of delivery.

Root Cause

Software design

Distribution Pattern

US - AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. OUS - Australia, Austria, Canada, Czech Republic, China, Croatia, Denmark, France, Finland, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom