Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591
FDA Recall Z-2500-2020 · Posted None
Advertisement
Recalling FirmXTANT Medical
Recall StatusTerminated
Product CodeHWA
Date Initiated2020-05-12
Date Terminated2021-03-19
Product Quantity4 units
Reason for Recall
The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.
Firm Action
On May 12, 2020 a customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to the three customers who received the affected devices.
On May 19, 2020, an additional customer letter (in an envelope marked URGENT) was Federal Expressed priority overnight-signature required to one of the three affected customers.
Root Cause
Process design
Distribution Pattern
AZ, TX and WV.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.