Open, Classified

Olympus Tracheal Intubation Fiberscope, Models LF-DP, LF-GP, & LF-TP.

FDA Recall Z-2504-2023 · Posted 2023-08-31

Recall StatusOpen, Classified
Product CodeEOQ
Date Initiated2023-07-27
Product Quantity10801 units

Reason for Recall

Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

Firm Action

Consignees were mailed an Urgent Medical Device Correction notification dated 7/27/2023 via UPS. The notification asks that consignees inspect their inventory for recalled devices and check clinical areas to see if any remain in use. Consignees are to ensure that all personnel, including clinical staff, are completely knowledgeable of this recall notification and addendum, which instructs users to slide the endotracheal tube along the entire length of the tracheal intubation scope insertion section to confirm compatibility before using on a patient. Consignees are asked to complete and return the provided response form to Sedgwick via email at [email protected] or by fax at 866-552-4917. Any questions about the acknowledgement form may be directed by phone to 866-737-4762. If product has been further distributed, please forward the notification to those secondary consignees and maintain records of your documentation process. Consignees are to report any complaints to Olympus Technical Assistance Center at 1-800-848-9024 (Option 1) and FDA Medwatch. Any further questions are to be directed to Tara Safi by email at [email protected].

Root Cause

Labeling Change Control

Distribution Pattern

US Nationwide distribution.