Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is designed to deliver stimulation to the hypoglossal nerve for the treatment of obstructive sleep apnea. The lead features a flexible, self-sizing cuff. Electrodes in the inner surface of the cuff deliver stimulation to the nerve. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
FDA Recall Z-2516-2026 · Posted 2026-06-18
Advertisement
Recalling FirmInspire Medical Systems Inc.
Recall StatusOpen, Classified
Product CodeMNQ
Date Initiated2026-05-11
Product Quantity10 impacted units
Reason for Recall
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Firm Action
Firm began notifying consignees on May 11, 2026. Letters were titled "URGENT Medical Device Correction."
The firm requests consignees' assistance in removing affected product from clinical inventory and returning them to Inspire.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Hong Kong.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.