Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014
FDA Recall Z-2519-2025 · Posted 2025-09-04
Advertisement
Recalling FirmLumicell, Inc.
Recall StatusOpen, Classified
Product CodeSAW
Date Initiated2025-07-31
Product Quantity200 units (20 packs containing 10 units per pack)
Reason for Recall
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
Firm Action
On July 31, 2025, URGENT: MEDICAL DECICE RECALL letters were sent to customers.
Actions to be Taken by Customer:
" Lumicell recommends that users dispose of the elastic bands and refrain from using
them during the procedure. The Lumicell DVS may be used without elastic bands.
" The elimination of elastic bands does not impact system performance. However, in
the absence of elastic bands, there is a possibility that the secondary Sterile Probe
Cover may shift during use. If this occurs, the Sterile Probe Cover may be
repositioned, and Lumicell DVS use may continue.
" If the included elastic bands are used, follow the Lumicell Direct Visualization
System (DVS) Instructions for Use to ensure the bands are counted before and after
the procedure to confirm the elastic bands are accounted for.
" Please forward and post this notice for awareness to those who use it. If device was
distributed or sold, please forward this notice to those accounts.
" Adverse reactions or quality problems experienced with the use of this product may
be reported to the FDA's MedWatch Adverse Event Reporting program either online,
by regular mail or by fax.
Please acknowledge receipt of this letter by completing the included response card.
Next Actions:
Lumicell is investigating the root cause of the issue and is working on a long-term solution.
Lumicell will contact you in the future regarding updates. If you have any questions or need
further assistance, please do not hesitate to reach out to our customer support team at 1-833-4-LUM-DVS (Monday through Friday, 9AM - 5PM Eastern Time) or email
[email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of CA, FL, MA, NC, TX.