Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator
FDA Recall Z-2525-2023 · Posted 2023-09-20
Advertisement
Recalling FirmHamilton Medical, Inc.
Recall StatusOpen, Classified
Product CodeCBK
Date Initiated2023-07-26
Product Quantity164 units
Reason for Recall
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Firm Action
On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions
1. Check for affected devices in your facility. Get in contact with your local
distribution partner and get the devices serviced with high priority.
2. Please sign the enclosed/attached sheet to confirm that, as an End-
Customer, you have received and duly noted this Medical Device
Correction.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.