Terminated

Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.

FDA Recall Z-2526-2018 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2018-05-25
Date Terminated2019-05-16
Product Quantity3 US

Reason for Recall

Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Firm Action

Siemens issued on May 25, 2018 an URGENT: MEDICAL DEVICE CORRECTION customer notification to affected customers and advised a hardware update will be initiated . Users advised of the problem, potential risks and action to be taken. Siemens will correct the error by replacing all affected cooling units with Update Instruction AX076/17/S. Users may continue to use the Artis system while awaiting implementation of the corrective action.

Root Cause

Process control

Distribution Pattern

HI , IL, MI, NC, NY, VA, WI

Official Source

View FDA Recall →