Artis Pheno. Image-Intensified Flouroscopic X-Ray System.
FDA Recall Z-2529-2025 · Posted 2025-09-05
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2025-08-12
Product Quantity20 units
Reason for Recall
Limited system movements after startup .
Firm Action
Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to [email protected] or by certified mail.
Root Cause
Software design
Distribution Pattern
US Nationwide distribution in the states of AL, AZ, FL, KY, MA, MI, NY, OR, A, TN, TX & UT.