Terminated

iConnect Access used with Ortho PACS software provides medical specialists with access to diagnostic quality images, reports, and various types of patient data over conventional TCP/IP (e.g., internet) networks.

FDA Recall Z-2532-2016 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-01-30
Date Terminated2021-05-07
Product Quantity146 sites may have the affected version for both products

Reason for Recall

Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.

Firm Action

The recalling firm issued a letter dated 1/28/2016 via e-mail on 1/31/2016 or via certified mail if they did not have an e-mail address for the customer.

Root Cause

Software design

Distribution Pattern

Distribution was made to medical facilities nationwide and to one foreign medical facility in New Zealand. There was no military or government distribution.